Spezifikationen
- Zustand
- neu
- Models sonimix 4001
- 6800 17 000 standard range6800 17 001 extended range
- Range
- From 10 to 1000 ppb in 1 ppb steps @ 3 L/minFrom 10ppb to 10 ppm @ 3 L/min + 10ppm to 35 ppm @ 0.9 L/min
- Dilution principle
- Sonic nozzles and high precision pressure regulators
- Dilution air
- Dry air from internal zero air (option) or external dry air source
- Generated gas flow
- 3000 ml/min for range 1900ml/min for range 2
- Accuracy
- 10 ppb to 100 ppb : +/- 2 ppb >100 ppb : 1.5% relative
- Ozone repeatability
- Better than 1% relative
- Warm up time
- < 45 min
- Stabilization time
- < 15 min
- O3 regulation
- Measure of the UV light absorption of the generated ozone by UV reference photometer. (US EPA method and Beer Lambert law). Including atmospheric pressure and room temperature corrections
- Dimensions
- 19cmx36cm, 50cm deep
- Net weight
- 15 Kg
- Uv lamp life time
- About 10’000 hours (with a mean concentration of 1 ppm)
- Life time
- Annual preventive maintenance
- Quality
- O3, NO and NO2 < 1 ppb – H2O < -40°C dew point
- Outlet port
- Teflon 1/4‘’ Swagelok fitting
- Power supply voltage
- 230V/50-60Hz – 117V/60Hz – 100V/60Hz
- Kategorie
- Luftreiniger in Italien
- Subkategorie
- Premium gas-kalibrationssysteme
- Inserat-ID
- 59850155
Beschreibung
The Sonimix 4001 is a stand alone ozone primary transfer standard able to generate ozone together with dry or wet air (optional) allowing to calibrate the ozone monitors in several points. It generates an outflow of 3 Nl/min air containing 10 to 1000 ppb O3, with stability and reproducibility of the concentration better than 1% relative. The extended range model generates 10 ppb to 10 ppm at 3 L/min and 10 ppm to 35 ppm at 0.9 L/min.The Sonimix 4001 is using pressure regulator and sonic nozzles and an ozone generator including thermostatisation and regulation by UV light photometer measurement with atmospheric pressure and temperature corrections (US EPA method of UV absorption by ozone, Beer-Lambert law). The ozone analyser is built with 1 reference UV light detectors that measure in alternance the UV intensity without any ozone and then the same UV light but through the generated ozone. With these 2 values, the processor calculates the amount of ozone generated.